Human Anti-MPV IgG (Anti-Monkeypox Virus IgG) ELISA Kit
Cat# NB-66-31989-48T
Size : 48T
Products
Overview
Standard curve
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Anti-MPV IgG and the recovery rates were calculated by comparing the measured value to the expected amount of Anti-MPV IgG in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-95% | 90% |
| EDTA plasma(n=5) | 79-93% | 86% |
| Heparin plasma(n=5) | 95-107% | 101% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Anti-MPV IgG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-102% | 91-99% | 83-95% | 89-97% |
| EDTA plasma(n=5) | 82-94% | 95-102% | 87-103% | 88-95% |
| Heparin plasma(n=5) | 87-95% | 78-91% | 97-103% | 95-101% |