Zebrafish 5-HT(5-Hydroxytryptamine) ELISA Kit
Cat# NB-66-23461-48T
Size : 48T
Products
Overview
Standard curve
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant 5-HT and the recovery rates were calculated by comparing the measured value to the expected amount of 5-HT in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-99% | 92% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 95-107% | 101% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 5-HT and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-102% | 87-95% | 85-96% | 92-102% |
| EDTA plasma(n=5) | 81-92% | 86-93% | 91-103% | 81-94% |
| Heparin plasma(n=5) | 86-98% | 92-101% | 95-104% | 87-96% |