Human CA19-9(Carbohydrate Antigen 19-9) ELISA Kit
Katalog-Nummer NB-66-31513-96Tx5
Size : 96Tx5
Products
Overview
Standard curve
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant CA19-9 and the recovery rates were calculated by comparing the measured value to the expected amount of CA19-9 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 90-103% | 96% |
| Heparin plasma(n=5) | 85-97% | 91% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CA19-9 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-97% | 95-106% | 83-98% | 93-105% |
| EDTA plasma(n=5) | 88-101% | 90-98% | 92-106% | 85-97% |
| Heparin plasma(n=5) | 91-104% | 82-96% | 88-101% | 90-99% |